SK-L200 SERIES is driver to communicate with temperature/humidity measuring instrument of SATO KEIRYOKI MFC. CO., LTD in Japan.
Importance) SATO SK-L200 SERIES device not demand data at Naver me computer, and it sends measured data at controller.
Therefore, you have to set as communication mode( Data sending mode ) by pressing CALL/OUT button after turn on the power of device.
<Figure 1> is read setting example of SATO SK-L200 SERIES communication driver.
![]() |
<Figure 1> Read setting example of SATO SK-L200 SERIES communication driver |
Notice) Device setting of SATO SK-L200 SERIES driver are fixed as 9600, 0, 8, 1.
Read schedule example)
READ | STATION | Read Command | Read Address (Don't care) | Save Address | Read Size ( Fixed to 1) |
READ, | 0, | S, | 0, | 0, | 1, |
1) STATION : Don't care ( Fixed to 0 )
2) Read Command : Don't care ( Fixed to 0 )
3) Read Start Address : Don't care ( Fixed to 0 )
4) Save Start Address for Communication Server : Saving start address of Communication Server. ( 0 ~ 9999 )
5) Read Size : Read size. ( Always set as 1, Refer to <Table 1> about real read data size and contents )
Memory Save Address | Contents | Remarks |
Save addr + 0 | Temperature x 10 | กษ |
Save addr + 1 | Humidity x 10 | % |
<Table 1> Contents and saveaddress which were read by S(reading entire data) reading command |
If you click the icon in protocol
option part at <Figure 1>, you can see the dialog box such as <Figure 2>. You
can also set read schedule by using this part.
![]() |
<Figure 2> Example of SATO SK-L200 SERIES driver's Option dialog box |
You can set read schedule by using ,
,
button and
listbox of <Figure 2>.
![]() |
<Figure 3> Example of SATO SK-L200 SERIES driver's read schedule Add/Edit dialog box |
There is no writing setting for SATO SK-L200 Series, because it is read only controller.
Serial communication cable connection
Connect Serial communication cable supported at SATO Co. to the bottom left terminal of SATO SK-L200 SERIES device. ( Refer to <Figure 4> )
![]() |
<Figure 4> Appearance of SATO SK-L200 SERIES device |
Connection of temperature/humidity sensor
Connect temperature/humidity sensor supported at SATO Co. to the upper left terminal of SATO SK-L200 SERIES device, such as <Figure 4>.
Power connection of device
Connect battery to the rear of SATO SK-L200 SERIES device or connect DC 6V power to the bottom left.
Device power ON/OFF
If you press POWER ON/OFF button of the front of SATO SK-L200 SERIES device between 1 ~ 2sec, you can ON/OFF the power.
Communication mode setting of device
If you press CALL/OUT button of the front of SATO SK-L200 SERIES device, you can set communication/non communication mode.
In case of communication mode, OUT letter is displayed at LCD screen.
Importance) To communicate with SATO device, device have to be set as communication mode( Data sending ).